A table of polyatomic interferences in ICP-MS
Links
- More Information:
- Download citation as: RIS | Dublin Core
Abstract
Spectroscopic interferences are probably the largest class of interferences in ICP-MS and are caused by atomic or molecular ions that have the same mass-to-charge as analytes of interest. Current ICP-MS instrumental software corrects for all known atomic “isobaric” interferences, or those caused by overlapping isotopes of different elements, but does not correct for most polyatomic interferences. Such interferences are caused by polyatomic ions that are formed from precursors having numerous sources, such as the sample matrix, reagents used for preparation, plasma gases, and entrained atmospheric gases.
Publication type | Article |
---|---|
Publication Subtype | Journal Article |
Title | A table of polyatomic interferences in ICP-MS |
Series title | Atomic Spectroscopy |
DOI | 10.46770/AS.1998.05.002 |
Volume | 19 |
Issue | 5 |
Year Published | 1998 |
Language | English |
Publisher | Atomic Spectroscopy Press |
Contributing office(s) | Columbia Environmental Research Center |
Description | 6 p. |
First page | 150 |
Last page | 155 |
Google Analytic Metrics | Metrics page |