Abstract

The δ2HVSMOW–SLAP value of total hydrogen of the international measurement standard NBS 22 oil was determined by a new method of sealing water in silver tubes for use in a thermal conversion elemental analysis (TC/EA) reduction unit. The isotopic fractionation of water due to evaporation is virtually non-existent in this silver-tube method. A new value for the δ2HVSMOW–SLAP of NBS 22 oil, calibrated with isotopic reference waters, was determined to be −116.9 ± 0.8‰ (1σ and n = 31).
Publication type Article
Publication Subtype Journal Article
Title Applying the silver-tube introduction method for thermal conversion elemental analyses and a new δ2H value for NBS 22 oil
Series title Rapid Communications in Mass Spectrometry
DOI 10.1002/rcm.4638
Volume 24
Issue 15
Year Published 2010
Language English
Publisher Wiley
Publisher location Hoboken, NJ
Description 8 p.
Larger Work Type Article
Larger Work Subtype Journal Article
Larger Work Title Rapid Communications in Mass Spectrometry
First page 2269
Last page 2276