Comprehensive inter-laboratory calibration of reference materials for d18O versus VSMOW using various on-line high-temperature conversion techniques

Rapid Communications in Mass Spectrometry

, , , , , , , , and
DOI: 10.1002/rcm.3958



"Internationally distributed organic and inorganic oxygen isotopic reference materials have been calibrated by six laboratories carrying out more than 5300 measurements using a variety of high-temperature conversion techniques (HTC) in an evaluation sponsored by the International Union of Pure and Applied Chemistry (IUPAC). To aid in the calibration of these reference materials, which span more than 125‰, an artificially enriched reference water (d18O of +78.91‰) and two barium sulfates (one depleted and one enriched in 18O) were prepared and calibrated relative to VSMOW2 and SLAP reference waters. These materials were used to calibrate the other isotopic reference materials in this study, which yielded: Reference material d18O and estimated combined uncertainty IAEA-602 benzoic acid +71.28?±?0.36‰ USGS35 sodium nitrate +56.81?±?0.31‰ IAEA-NO-3 potassium nitrate +25.32?±?0.29‰ IAEA-601 benzoic acid +23.14?±?0.19‰ IAEA-SO-5 barium sulfate +12.13?±?0.33‰ NBS 127 barium sulfate +8.59?±?0.26‰ VSMOW2 water 0‰ IAEA-600 caffeine -3.48?±?0.53‰ IAEA-SO-6 barium sulfate -11.35?±?0.31‰ USGS34 potassium nitrate -27.78?±?0.37‰ SLAP water -55.5‰ The seemingly large estimated combined uncertainties arise from differences in instrumentation and methodology and difficulty in accounting for all measurement bias. They are composed of the 3-fold standard errors directly calculated from the measurements and provision for systematic errors discussed in this paper. A primary conclusion of this study is that nitrate samples analyzed for d18O should be analyzed with internationally distributed isotopic nitrates, and likewise for sulfates and organics. Authors reporting relative differences of oxygen-isotope ratios (d18O) of nitrates, sulfates, or organic material should explicitly state in their reports the d18O values of two or more internationally distributed nitrates (USGS34, IAEA-NO-3, and USGS35), sulfates (IAEA-SO-5, IAEA-SO-6, and NBS 127), or organic material (IAEA-601 benzoic acid, IAEA-602 benzoic acid, and IAEA-600 caffeine), as appropriate to the material being analyzed, had these reference materials been analyzed with unknowns. This procedure ensures that readers will be able to normalize the d18O values at a later time should it become necessary. The high-temperature reduction technique for analyzing d18O and d2H is not as widely applicable as the well-established combustion technique for carbon and nitrogen stable isotope determination. To obtain the most reliable stable isotope data, materials should be treated in an identical fashion; within the same sequence of analyses, samples should be compared with working reference materials that are as similar in nature and in isotopic composition as feasible."

Additional publication details

Publication type:
Publication Subtype:
Journal Article
Comprehensive inter-laboratory calibration of reference materials for d18O versus VSMOW using various on-line high-temperature conversion techniques
Series title:
Rapid Communications in Mass Spectrometry
Year Published:
Contributing office(s):
Branch of Regional Research-Eastern Region
Larger Work Type:
Larger Work Subtype:
Journal Article
Larger Work Title:
Rapid Communications in Mass Spectrometry
First page:
Last page:
United States