| Abstract: | As delivered from Finnigan MAT, the emission current of the Leybold-Heraeus model IM210 ionization vacuum gage on a model 251 isotope ratio mass spectrometer can turn on as soon as the ‘ PUMP ON ‘ switch is depressed, before a good vacuum in the mass spectrometer is attained. The filament in the source of the mass spectrometer may turn on at the same time, thereby shortening its life or burning it out if the vacuum is poor. This design flaw can be corrected by a simple modification of the electronic circuitry. (Author ‘s abstract) |
| Genre: | USGS Numbered Series |
| ProdID: | 16542 |
| Citation Author: | Wildman, J. D. |
| Citation Contributing Office: | |
| Citation Datum: | |
| Citation Day: | |
| Citation Edition: | - |
| Citation Editor: | |
| Citation End Page: | |
| Citation Issue: | |
| Citation Keywords: | |
| Citation Language: | ENGLISH |
| Citation Larger Work Title: | |
| Citation LatN: | |
| Citation LatS: | |
| Citation LonE: | |
| Citation LonW: | |
| Citation Month: | |
| Citation No Pagination: | |
| Citation Number Of Pages: | |
| Citation Online Only Flag: | |
| Citation Phsyical Description: | 4 leaves ;28 cm. |
| Citation Projection: | |
| Citation Public Comments: | |
| Citation Publisher: | U.S. Geological Survey, |
| Citation Series: | Open-File Report |
| Citation Series Code: | OFR |
| Citation Series Number: | 86-484 |
| Citation Search Results Text: | Improvement of filament-protection circuitry in a Finnigan model 251 mass spectrometer; 1986; OFR; 86-484; Wildman, J. D. |
| Citation Start Page: | |
| Citation Volume: | |
| Citation Year: | 1986 |
| Type: | citation/reference |
| Text: | Improvement of filament-protection circuitry in a Finnigan model 251 mass spectrometer; 1986; OFR; 86-484; Wildman, J. D. |
| URL (THUMBNAIL): | http://pubs.usgs.gov/of/1986/0484/report-thumb.jpg |
| URL (DOCUMENT): | http://pubs.usgs.gov/of/1986/0484/report.pdf |
| Date Other: | Sat, 1 Jan 1994 00:00 -0600 |
| Publisher: | U.S. Geological Survey, |