| Abstract: | INTRODUCTION
This report documents the results of a series of experiments conducted by the authors at the Albuquerque Seismological Laboratory (ASl) during the spring and summer of 1991; the object of these experiments was to obtain and document quantitative performance comparisons of three methods of installing STS-1 seismometers.
Historically, ASL has installed STS-1 sensors by cementing their thick glass base plates to the concrete floor of the vault (see Peterson and Tilgner, 1985, p 44 and Figure 31, p 51 for the details of this installation technique). This installation technique proved to be fairly satisfactory for the China Digital Seismic Network and for several sets of STS-1 sensors installed in other locations since that time. However, the cementing operation is rather labor intensive and the concrete requires a lengthy (about 1 week) curing time during which the sensor installed on it is noisy. In addition it is difficult to assure that all air bubbles have been removed from the interface between the cement and the glass base plate. If air bubbles are present beneath the plate, horizontal sensors can be unacceptably noisy. Moving a sensor installed in this manner requires the purchase of a new glass base plate because the old plate normally can not be removed without breakage.
Therefore, this study was undertaken with the aim of developing an improved method of installing STS-1‘s. The goals were to develop a method which requires less field site labor during the installation and assures a higher quality installation when finished. In addition, the improved installation technique should promote portability.
Two alternate installation techniques were evaluated in this study. One method replaces the cement between the base plate and the vault floor with sand. This method has been used in the French Geoscope program and in several IRIS/IDA installations made by the University of California at San Diego (UCSD) and possibly others. It is easily implemented in the field and is quite cheap. The other method utilizes a so called warpless housing designed by E. Wielandt and implemented at ASL. This housing is quite similar to the case design of the STS-2 sensor system. It is designed to minimize the effects of atmospheric pressure variations on the sealed housing. |
| Genre: | USGS Numbered Series |
| ProdID: | 31534 |
| Citation Author: | Holcomb, L. Gary; Hutt, Charles R. |
| Citation Contributing Office: | USGS Albuquerque Seismological Laboratory |
| Citation Datum: | |
| Citation Day: | |
| Citation Edition: | - |
| Citation Editor: | |
| Citation End Page: | |
| Citation Issue: | |
| Citation Keywords: | |
| Citation Language: | ENGLISH |
| Citation Larger Work Title: | |
| Citation LatN: | |
| Citation LatS: | |
| Citation LonE: | |
| Citation LonW: | |
| Citation Month: | |
| Citation No Pagination: | |
| Citation Number Of Pages: | |
| Citation Online Only Flag: | |
| Citation Phsyical Description: | 37 p. |
| Citation Projection: | |
| Citation Public Comments: | |
| Citation Publisher: | Geological Survey (U.S.) |
| Citation Series: | Open-File Report |
| Citation Series Code: | OFR |
| Citation Series Number: | 92-302 |
| Citation Search Results Text: | An Evaluation of Installation Methods for STS-1 Seismometers; 1992; OFR; 92-302; Holcomb, L. Gary; Hutt, Charles R. |
| Citation Start Page: | |
| Citation Volume: | |
| Citation Year: | 1992 |
| Type: | citation/reference |
| Text: | An Evaluation of Installation Methods for STS-1 Seismometers; 1992; OFR; 92-302; Holcomb, L. Gary; Hutt, Charles R. |
| URL (THUMBNAIL): | http://pubs.er.usgs.gov/thumbnails/usgs_thumb.jpg |
| URL (DOCUMENT): | http://earthquake.usgs.gov/regional/asl/pubs/files/ofr92-302.pdf |
| Date Other: | Mon, 1 Apr 2002 00:00 -0600 |
| Publisher: | Geological Survey (U.S.) |