The precision and accuracy of iterative and non-iterative methods of photopeak integration in activation analysis, with particular reference to the analysis of multiplets

Journal of Radioanalytical Chemistry
By:

Links

Abstract

The relative precisions obtainable using two digital methods, and three iterative least squares fitting procedures of photopeak integration have been compared empirically using 12 replicate counts of a test sample with 14 photopeaks of varying intensity. The accuracy by which the various iterative fitting methods could analyse synthetic doublets has also been evaluated, and compared with a simple non-iterative approach. ?? 1977 Akade??miai Kiado??.