Improved spectrophotometric cell for hydrothermal solutions

Review of Scientific Instruments
By: , and 

Links

Abstract

A simple, inexpensive spectrophotometric cell was designed for use with aqueous solutions for which temperature is a maximum of 325??C and pressure, 28 MPa. The cell has an internal volume of 5 ml and a path length of 1.31 cm. Each furnace assembly is 120 mm in diameter ?? 150 mm high and will fit into most commercial spectrophotometers. Temperature is controlled by a standard set-point controller and a balancing circuit that is used to maintain the temperature of the sample and reference cell within 1??C of each other at any temperature.
Publication type Article
Publication Subtype Journal Article
Title Improved spectrophotometric cell for hydrothermal solutions
Series title Review of Scientific Instruments
DOI 10.1063/1.1136597
Volume 52
Issue 3
Year Published 1981
Language English
Larger Work Type Article
Larger Work Subtype Journal Article
Larger Work Title Review of Scientific Instruments
First page 428
Last page 431
Google Analytic Metrics Metrics page
Additional publication details