Use of field-applied quality control samples to monitor performance of a Goulden large-sample extractor/GC-MS method for pesticides in water

International Journal of Environmental Analytical Chemistry
By: , and 

Links

  • The Publications Warehouse does not have links to digital versions of this publication at this time
  • Download citation as: RIS | Dublin Core

Abstract

Since 1985, the Goulden large-sample extractor (GLSE) has been used to isolate a broad array of trace-organic contaminants from large volumes of water. In this study, field-applied quality control measures, including matrix and surrogate spikes and blanks, were used to monitor method performance from GLSE extraction through GC-MS analysis. The method was applied to the determination of multiple classes of pesticides isolated from 4- to 112-L filtered surface-water samples. Average recoveries of six surrogate compounds ranged from 84 ?? 18% for [2H10]diazinon to 15 ?? 13% for 4,4'-[2H8]DDT, the low recoveries for which were largely a result of unmonitored breakdown of this surrogate by the GC injection system. Field-matrix-spike samples were prepared by fortifying 10-L, 35-L, and 110-L filtered surface-water samples with 68 pesticides to amended concentrations of 11- to 50-ng/L each. Recoveries ranged from not detected to greater than 100%. Variability in pesticide recoveries from triplicate 10-L water samples collected at one site averaged 5.7% relative standard deviation and did not exceed 19%.Since 1985, the Goulden large-sample extractor (GLSE) has been used to isolate a broad array of trace-organic contaminants from large volumes of water. In this study, field-applied quality control measures, including matrix and surrogate spikes and blanks, were used to monitor method performance from GLSE extraction through GC-MS analysis. The method was applied to the determination of multiple classes of pesticides isolated from 4- to 112-L filtered surface-water samples. Average recoveries of six surrogate compounds ranged from 84 ?? 18% for [2H10]diazinon to 15 ?? 13% for 4,4???-[2H8]DDT, the low recoveries for which were largely a result of unmonitored breakdown of this surrogate by the GC injection system. Field-matrix-spike samples were prepared by fortifying 10-L, 35-L, and 110-L filtered surface-water samples with 68 pesticides to amended concentrations of 11- to 50-ng/L each. Recoveries ranged from not detected to greater than 100%. Variability in pesticide recoveries from triplicate 10-L water samples collected at one site averaged 5.7% relative standard deviation and did not exceed 19%.

Additional publication details

Publication type Article
Publication Subtype Journal Article
Title Use of field-applied quality control samples to monitor performance of a Goulden large-sample extractor/GC-MS method for pesticides in water
Series title International Journal of Environmental Analytical Chemistry
Volume 77
Issue 1
Year Published 2000
Language English
Publisher Gordon & Breach Science Publ Inc
Publisher location Newark, NJ, United States
Larger Work Type Article
Larger Work Subtype Journal Article
Larger Work Title International Journal of Environmental Analytical Chemistry
First page 39
Last page 62