Laboratory evaluation of the Level TROLL 100 manufactured by In-Situ Inc.: results of pressure and temperature tests

Open-File Report 2013-1173
By: , and 

Links

Abstract

The Level TROLL 100 manufactured by In-Situ Inc. was evaluated by the U.S. Geological Survey (USGS) Hydrologic Instrumentation Facility (HIF) for conformance to the manufacturer’s accuracy specifications for measuring pressure throughout the device’s operating temperature range. The Level TROLL 100 is a submersible, sealed, water-level sensing device with an operating pressure range equivalent to 0 to 30 feet of water over a temperature range of −20 to 50 degrees Celsius (°C). The device met the manufacturer’s stated accuracy specifications for pressure within its temperature-compensated operating range of 0 to 50 °C. The device’s accuracy specifications did not meet established USGS requirements for primary water-stage sensors used in the operation of streamgages, but the Level TROLL 100 may be suitable for other hydrologic data-collection applications. As a note, the Level TROLL 100 is not designed to meet USGS accuracy requirements. Manufacturer accuracy specifications were evaluated, and the procedures followed and the results obtained are described in this report. USGS accuracy requirements are routinely examined and reported when instruments are evaluated at the HIF.
Publication type Report
Publication Subtype USGS Numbered Series
Title Laboratory evaluation of the Level TROLL 100 manufactured by In-Situ Inc.: results of pressure and temperature tests
Series title Open-File Report
Series number 2013-1173
DOI 10.3133/ofr20131173
Year Published 2013
Language English
Publisher U.S. Geological Survey
Publisher location Reston, VA
Contributing office(s) Hydrologic Instrumentation Facility
Description v, 12 p.
Online Only (Y/N) Y
Google Analytic Metrics Metrics page
Additional publication details