Rapid and precise determination of ATP using a modified photometer

Open-File Report 80-1194
By:  and 

Links

Abstract

An inexpensive delay timer was designed to modify a commercially available ATP photometer which allows a disposable tip pipette to be used for injecting either enzyme or sample into the reaction cuvette. The disposable tip pipette is as precise and accurate as a fixed-needle syringe but eliminates the problem of sample contamination and decreases analytical time. (USGS)
Publication type Report
Publication Subtype USGS Numbered Series
Title Rapid and precise determination of ATP using a modified photometer
Series title Open-File Report
Series number 80-1194
DOI 10.3133/ofr801194
Year Published 1980
Language English
Publisher U.S. Geological Survey
Publisher location Reston, VA
Description iii, 10 p.
Google Analytic Metrics Metrics page
Additional publication details