Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of antimony by automated-hydride atomic absorption spectrophotometry

Open-File Report 93-664
By:  and 

Links

Abstract

The analysis of natural-water samples for antimony by automated-hydride atomic absorption spectrophotometry is described. Samples are prepared for analysis by addition of potassium and hydrochloric acid followed by an autoclave digestion. After the digestion, potassium iodide and sodium borohydride are added automatically. Antimony hydride (stibine) gas is generated, then swept into a heated quartz cell for determination of antimony by atomic absorption spectrophotometry. Precision and accuracy data are presented. Results obtained on standard reference water samples agree with means established by interlaboratory studies. Spike recoveries for actual samples range from 90 to 114 percent. Replicate analyses of water samples of varying matrices give relative standard deviations from 3 to 10 percent.
Publication type Report
Publication Subtype USGS Numbered Series
Title Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory-Determination of antimony by automated-hydride atomic absorption spectrophotometry
Series title Open-File Report
Series number 93-664
DOI 10.3133/ofr93664
Year Published 1994
Language English
Publisher U.S. Geological Survey : Books and Open-File Reports Section [distributor],
Contributing office(s) National Water Quality Laboratory
Description iv, 17 p.
Google Analytic Metrics Metrics page
Additional publication details