Portable precision dc voltage-current transfer standard for electrometer calibration

Review of Scientific Instruments
By:  and 

Links

Abstract

A circuit design is presented for an instrument providing a highly stable and fully adjustable voltage and current in the range of 0-1.999 V or 0-199.9 mV and 10-11-10-15 A. This instrument is used to verify the calibration and performance of dc and vibrating reed electrometers and chart recorders on mass spectrometers of the USGS Isotope Laboratories in Denver.
Publication type Article
Publication Subtype Journal Article
Title Portable precision dc voltage-current transfer standard for electrometer calibration
Series title Review of Scientific Instruments
DOI 10.1063/1.1137127
Volume 53
Issue 8
Year Published 1982
Language English
Larger Work Type Article
Larger Work Subtype Journal Article
Larger Work Title Review of Scientific Instruments
First page 1290
Last page 1291
Google Analytic Metrics Metrics page
Additional publication details