thumbnail

Determination of ammonium in a buddingtonite sample by ion-chromatography

Talanta
By:  and 

Links

  • The Publications Warehouse does not have links to digital versions of this publication at this time
  • Download citation as: RIS | Dublin Core

Abstract

An ion-chromatographic method for the direct determination of ammonium, potassium, and sodium in geologic materials is described. Samples are decomposed with a mixture of hydrofluoric and hydrochloric acids in a sealed polycarbonate bottle heated in a microwave oven. The ion-chromatograph separates the cations and determines them by conductivity measurement. The ammonium concentrations thus determined have been verified by use of an ammonia-specific electrode. A total of 32 analyses of ammonium salts by both techniques showed an average error of -4%, with a relative standard deviation (RSD) of 6%. The ammonium concentrations found in a buddingtonite sample had an RSD of 2.2% and their mean agreed with that obtained by the Kjeldahl method. By use of the prescribed dilution of the sample, detection limits of 0.1% can be achieved for all three cations. ?? 1986.
Publication type Article
Publication Subtype Journal Article
Title Determination of ammonium in a buddingtonite sample by ion-chromatography
Series title Talanta
Volume 33
Issue 6
Year Published 1986
Language English
Larger Work Type Article
Larger Work Subtype Journal Article
Larger Work Title Talanta
First page 495
Last page 498
Google Analytic Metrics Metrics page
Additional publication details